Chapter 8 _____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 173

Dewpoint Temperature (HMT337 Warmed
Probe Option)

Find the intersection of the dewpoint temperature curve and the dewpoint
difference reading (process temperature-dewpoint temperature) on the x-
axis and read the accuracy in dewpoint measurement on the y-axis.

0508-017

Figure 81

Accuracy in Dewpoint Measurement

Operating Environment

Operating temperature range

for humidity measurement

-70 ... +180 °C (-94 ... +356 °F)
see probe specifications

for transmitter body electronics

-40 ... +60 °C (40 ... +140 °F)

with display

0 ... +60 °C (+32 ... +140 °F)


Storage temperature range

without display

-55 … +80 °C (-67 … +176 °F)

with display

-40 … +80 °C (-40 … +176 °F)


Electromagnetic compatibility

EN61326-1: Electrical equipment for
measurement, control, and laboratory use –
EMC requirements – for use in industrial
locations