Testing operation of relays, Rtest – Vaisala HMT330 User Manual

Page 134

Advertising
background image

User's Guide _______________________________________________________________________

132 __________________________________________________________________ M210566EN-K

Testing Operation of Relays

Testing activates relays even if they are disabled.

Use the module push buttons to activate the relays. Press the REL 1 or
REL 2 button to activate the corresponding relay.

Relay is activated:

led is lit

Relay is not activated:

led is not lit

Use the display/keypad to test the operation of relays.

1.

Open the Main Menu by pressing any of the arrow buttons.

2.

Select System

, press the ►arrow button.

3.

Select Diagnostics

, press the ►arrow button.

4.

Select Relay tests

, press the ►arrow button.

5.

Select Invert relay 1... , press TEST. Now the selected relay
output is forced to opposite state. Press OK to return to normal
operation.

6.

Press EXIT to return to the basic display.

RTEST

Use the serial line command RTEST to test the operation of the relays.

RTEST [x1 x2 x3 x4]<cr>

where

x

=

ON/OFF

Example: Activate and then release all four relays.

>rtest on on on on
ON ON ON ON
>
>rtest off off off off
OFF OFF OFF OFF
>

Enter the command RTEST without parameters to stop testing.

Advertising