Figure 77, Adjustment and purge buttons, Figure 78 – Vaisala HMT330 User Manual

Page 158: Adjustment menu, Table 34, Indicator led functions

Advertising
background image

User's Guide _______________________________________________________________________

156 __________________________________________________________________ M210566EN-K

0508-013

Figure 77

Adjustment and Purge Buttons

The following numbers refer to Figure 77 above:

1

=

Indicator LED

2

=

Adjustment button

3

=

Press the purge buttons simultaneously to start chemical
purge (if available)

Adjustment menu is displayed only when ADJ button (on the
motherboard inside the transmitter) is pressed.

0706-007

Figure 78

Adjustment Menu

Table 34

Indicator Led Functions

Indicator Led Function

Description

LED off

adjustment locked

LED on

adjustment available

LED blinking evenly

measurement not stabilized

LED blinking with short pulses

performing chemical purge

NOTE

If using a warmed probe (HMT337 option), probe heating will be
interrupted when ADJ key is pressed. Allow sufficient time for the
probe to reach ambient temperature before starting the adjustment
procedure.

NOTE

Fixed pressure compensation value of 1013.25 hPa is used when in
adjustment mode

Advertising