Chapter 8 _____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 169

CHAPTER 8

TECHNICAL DATA

This chapter provides the technical data of the product.

Specifications

Performance

Relative Humidity

Measurement range

0 ... 100 %RH

Accuracy (including non-linearity, hysteresis and repeatability)

with

HUMICAP®180

for typical applications

HUMICAP®180R

for typical applications

HUMICAP®180C

for applications with chemical purge and/or
warmed probe

HUMICAP®180RC

for applications with chemical purge and/or
warmed probe

at +15 ... 25 °C

± 1 %RH (0 ... 90 %RH)

± 1.7 %RH (90 ... 100 %RH)

at -20 ... +40 °C

± (1.0 + 0.008 × reading) %RH

at -40 ... + 180 °C

± (1.5 + 0.015 × reading) %RH


Factory calibration uncertainty (+20 °C)

±0.6 %RH (0 ... 40 %RH)

±1.0 %RH (40 ... 97 %RH)

(Defined as ± 2 standard deviation limits.
Small variations possible, see also calibration
certificate.)

Response time (90 %) for HUMICAP®180 and HUMICAP

®

180C in still air

8 s with grid filter

20 s with grid + steel netting filter

40 s with sintered filter


Response time (90 %) for HUMICAP®180R and HUMICAP®180RC
at 20 °C in 0.1 m/s air flow

17 s with grid filter

50 s with grid + steel netting filter

60 s with sintered filter