6 replacement parts, Replacement parts, 13. disk filter of early ap200 intake assembly – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 22: 14. single desiccant pack

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

CR1000 in the AP200. The apps support field maintenance tasks such as

viewing and collecting data, setting the clock, and downloading programs.

4.1.6 Replacement Parts

Syringe Filter: Intake assemblies shipped after August 2013 use a syringe

filter with Luer lock connections. The filter is an in-line 2.5-cm (1.0-in)

diameter, PTFE membrane filter (FIGURE 4-12) of 3-micron pore size. It is

used to replace dirty filter elements in the AP200 intake assembly and is

available as pn 29998.

FIGURE 4-12. Syringe filter of current AP200 intake assembly

Disk Filter: Intake assemblies shipped prior to August 2013 used a different

filter, pn 27809. This filter is a 2.5-cm (1.0-in) diameter, sintered stainless

steel disk filter (FIGURE 4-13) of 10-micron pore size. It is used to replace

dirty filter elements in the AP200 intake assembly.

FIGURE 4-13. Disk filter of early AP200 intake assembly

Silica Desiccant Bags: The 4905 Single 4-unit Silica Desiccant Bag (FIGURE

4-14) is used to desiccate the AP200 system enclosure and should be

periodically replaced. The 4905 can be purchased in quantities of 20 as pn

6714.

FIGURE 4-14. Single desiccant pack