4.3.3 Intake Assembly

Dimensions:

27.9 cm x 12.5 cm x 19.0 cm (11.0 in x 5.0

in x 7.5 in)

Weight:

1.4 kg (3.1 lb)

Filter:

1.0 in diameter, inline PTFE membrane

filter, 3 micron pore size, CSI pn 29998

Orifice:

0.007-in inside diameter

Heater:

560 ohms (0.25 W at 12Vdc)

Mixing volume:

750 ml

Sample connection:

0.25-in Swagelok

®

Heater cable entry seals

Number of connections:

3 (one in, and up to two out)

Cable diameter:

2.8 mm to 6.6 mm (0.11 in to 0.26 in)

Heater cable screw terminals

Wire diameter:

26 to 12 AWG

Wire stripping length:

5.0 mm (0.2 in)

Screw tightening torque: 0.4 Newton meter (N

m)

5. Installation

The following tools are required to install the AP200 system in the field.

Additional tools may be required for a user-supplied tripod or tower.

9/16-in open-end wrench

Adjustable wrench

Tubing cutter (included with AP200)

Small, flat-tip screwdriver (included with AP200)

Wire stripping tool

5.1 Mounting

5.1.1 Support Structure

The AP200 system has two types of components that must be mounted to a

support structure: the intake assemblies and the AP200 system enclosure. The

support structure itself is not included in the AP200 so that it can be tailored to

specific needs, but several options are available. Contact a Campbell Scientific

application engineer for more information on site-specific mounting options.

The following installation notes show a typical application using a UT30

tower.

The AP200 intake assemblies are designed to mount on a vertical pipe of 1.3-

cm to 5.1-cm (0.5-in to 2.0-in) diameter. They should be mounted at the

positions where the air is to be sampled.

The AP200 system enclosure should be mounted where it can be accessed

easily to retrieve data from the CF cards on the datalogger. The AP200 can be

configured with one of these mounting options, similar to the standard

ENC16/18 enclosure:

Triangular tower (UT10, UT20, or UT30)

Tripod mast (1.5-in to 1.9-in diameter)

Tripod leg (CM106 or CM106K tripod only)

20