3 h2o span, O span, And co – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 38

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

FIGURE 5-8. AP200 system enclosure configured with cylinders of

zero air and CO

2

Flow meters and needle valves are not needed because the

AP200 zero and span inlets have flow restriction inside the

system enclosure to set the flow rate.

Make sure there are no leaks in the regulators or the connections

to the valve module. For automatic operation, the tank shutoff

valves are left continuously open. A plumbing leak could cause

the contents of the tank to be lost.

When inlets are not in use, replace the Swagelok

®

plug to keep

the system clean.

Open the shut-off valves on the cylinders and set the pressure regulators for 0-

psig delivery pressure.

If the pressure is inadvertently adjusted to high, the tube fitting

may need to be slightly loosened to bleed off the excess pressure.

Retighten the fitting when the proper setting is reached.

5.2.3 H

2

O Span

The H

2

O span can be performed only as a manual operation. Automated H

2

O

span is not feasible because it would require a dewpoint generator to provide

the H

2

O span gas all the time. The H

2

O span inlet is not bypass equipped, so it

will flow only when selected. This inlet can be connected to the output of a

dewpoint generator using a tee at the inlet, as shown in FIGURE 5-9, to bleed

off excess flow and avoid pressurizing the dewpoint generator.

NOTE

NOTE

26