Appendix C. Public Variables

Some of the variables in the AP200’s CRBasic program are included in the

Public table. These public variables may be displayed or edited with a

keyboard display or PC. Other program variables are hidden from the user to

reduce clutter in the Public table. Many of these public variables are saved in

the output tables. Some of the public variables allow the user to set the

operation of the system or to give diagnostic information. The intended usage

of the public variables is categorized as follows:

IRGA

measured directly by the IRGA

TAIR

air temperature (defined only if

N_AirTemps > 0)

DIAG

diagnostic

USER

user setting

INFO

provides information on system status

INFO/USER conditional user setting (may set if the sequence is stopped)

CONFIG

system configuration parameter - saved in file

sys_conf_var.dat.

The public variables are listed in TABLE C-1.

TABLE C-1. Public Variables

Usage

Variable Name

Units

Description

RecNum

RN

record number

TimeStamp

TS

date and time the record was measured

DIAG

diag_AP200

AP200 diagnostic word (nonzero value indicates a

problem detected)

INFO

interval_counter

number of scans since the start of this output interval (half

hour)

IRGA

CO2

ppm

CO

2

concentration measured by the IRGA

IRGA

H2O

ppt

H

2

O concentration measured by the IRGA

IRGA

cell_tmpr

°C

IRGA sample cell temperature

IRGA

cell_press

kPa

IRGA sample cell pressure

IRGA

Td_cell

°C

dewpoint temperature measured by the IRGA

INFO

Td_ambient

°C

Td_cell, corrected to ambient pressure

DIAG

sample_flow

ml·min

-1

sample flow to the IRGA

USER

STARTsequence

set = to True to start the automatic valve sequence

USER

STOPsequence

set = to True to stop the automatic valve sequence

C-1