To remove the IRGA from the AP200:

1. Disconnect power from the AP200.

2. Disconnect the power connector from the IRGA. Leave the wires

connected to the terminal block and pull the terminal block out of the

IRGA. If the IRGA is to be powered outside the AP200, use the spare

terminal block supplied with the IRGA.

3. Disconnect the serial cable from the IRGA.

4. Disconnect the inlet and outlet tubes from the IRGA. Hold the tube to

prevent it from rotating while loosening the knurled nut. Pull the tube off

the fitting.

5. To keep the AP200 and the IRGA clean, connect the “

IN” tube to the

OUT” tube using the union fitting in the storage pocket in the door of the

AP200 system enclosure. Plug the IRGA fittings with the green, vinyl

caps.

6. Undo the Velcro strap and remove the IRGA from the mounting bracket.

FIGURE 7-5. Components and fittings of IRGA installation

7.4.2 Configuring the IRGA

The IRGA is automatically configured by the AP200 program. This

configuration is sent to the IRGA when the program compiles or if the IRGA

fails to respond for five consecutive samples. Messages are written to the