2 do co2 and h2o zero, Do co, And h – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 54: O zero, 2 do co

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

If

seq_ACTIVE is True, then set STOPsequence = True to stop the sequence

and then verify that

seq_ACTIVE changes to False.

Set

valve_number to Zero (9). Look at the LEDs on the valve module to

confirm the selected valve is now active.

The

CO2_SPAN_PPM variable displays the concentration in the CO

2

span

tank.

Check the value of

diag_AP200. If it is not zero, a problem in the AP200

system should be resolved before continuing (see Appendix B, AP200

Diagnostics).

Wait for the values of CO

2

and H

2

O to stabilize. Normally, this takes one or

two minutes.

Record the value of CO

2

. This is the measured CO

2

concentration in ppm. If

the zero cylinder is selected, the value should be close to zero. If the CO

2

span

tank is selected, it should be close to the value of

CO2_SPAN_PPM.

Record the value of H

2

O. This is the measured H

2

O concentration in ppt. The

value should be close to zero for either cylinder.

Set

valve_number to CO

2

span (10) and repeat the previous steps.

Press

<Esc> to return to the Manual Zero/Span menu.

6.2.2 Do CO

2

and H

2

O Zero

At the

Manual Zero/Span menu, select <Do Zero (CO2&H2O)>. This menu

guides the user through the steps to flow the zero gas and to send the command

to zero the IRGA.

If

seq_ACTIVE is True, then set STOPsequence = True to stop the

sequence. Then verify

seq_ACTIVE changes to False.

Set

valve_number to Zero (9). Look at the LEDs on the valve module to

confirm the Zero valve is now active.

Check the value of

diag_AP200. If it is not zero, a problem in the AP200

system should be resolved before continuing (see Appendix B, AP200

Diagnostics).

42