If

seq_ACTIVE is True, then set STOPsequence = True to stop the

sequence, and then verify

seq_ACTIVE changes to False.

Variable

cell_press is the pressure in the IRGA sample cell. Set pump_ON =

False to turn the pump off and it should be possible to hear the pump stop.

Wait for the value of

cell_press to rise and stabilize to ambient pressure. This

may take several minutes as the intake tubes and mixing volumes fill with air.

Make sure the dewpoint generator is connected to the H

2

O Span inlet as shown

in Section 5.2.3, H

2

O Span. Consult the dewpoint generator’s user manual for

proper operation of the dewpoint generator.

Enter the dewpoint generator’s temperature setpoint (in °C) for

H2OSpanDewPt.

While the pump is off and the IRGA sample cell is still at ambient pressure, set

SETpressAmbient = True. The AP200 will store the value of cell_press in

PRESS_AMBIENT. Verify that the value in PRESS_AMBIENT now

matches

cell_press.

The next two variables are

Td_cell and Td_ambient. Td_cell is the dewpoint

temperature measured by the IRGA.

Td_ambient is corrected for the

difference in pressure between the sample cell and ambient (see Appendix I,

Useful Equations). With the pump off, these two dewpoint temperatures

should match.

Step down to

pump_ON and set it to True to turn the pump on. It should be

possible to hear the pump start to run at full speed and then slow down to

stabilize the pump pressure to its setpoint.

While the pump is on and the IRGA sample cell is stable at its normal

operating pressure, set

SETspanCellP = True. The AP200 will store the value

of

cell_press in SpanCellP. Verify that the value in SpanCellP now matches

that of

cell_press.

Set

valve_number to H2Ospan (11). Look at the LEDs on the valve module

to confirm the H

2

O span valve is now active.

Check the value of

diag_AP200. If it is not zero, a problem in the AP200

system should be resolved before continuing (see Appendix B, AP200

Diagnostics).

Wait for the value of

Td_ambient to stabilize. This normally takes one to two

minutes.

Set

DO_H2O_span to True. The AP200 will send the command to the IRGA

to do an H

2

O span. While the IRGA is setting the span, the value for

Td_ambient will not be available and NAN will be displayed. This will take

approximately 10 s. When the process is complete,

Td_ambient will again be

displayed. Verify that the value is close to the value of

H2OSpanDewPt.