Appendix C. Public Variables

TABLE C-1. Public Variables

Usage

Variable Name

Units

Description

TAIR

T_air(5)

°C

air temperature profile (optional, defined only if

N_AirTemps > 4)

TAIR

T_air(6)

°C

air temperature profile (optional, defined only if

N_AirTemps > 5)

TAIR

T_air(7)

°C

air temperature profile (optional, defined only if

N_AirTemps > 6)

TAIR

T_air(8)

°C

air temperature profile (optional, defined only if

N_AirTemps > 7)

USER

H2OSpanDewPt

°C

setpoint temperature on dewpoint generator; used for H

2

O

span

USER

SETpressAmbient

set = to True for AP200 to store the value of cell_press in

PressAmbient

USER

SETspanCellP

Set = True for AP200 to store the value of cell_press in

SpanCellP

USER

SpanCellP

kPa

sample cell operating pressure; used for H

2

O span

INFO/USER DO_zero

set = to True for AP200 to send CO

2

& H

2

O zero

commands to the IRGA

INFO/USER DO_CO2_span

set = to True for AP200 to send CO

2

span command to the

IRGA

INFO/USER DO_H2O_span

Set = True for AP200 to send H

2

O span command to the

IRGA

CONFIG

PRESS_AMBIENT

kPa

ambient pressure; used to calculate Td_ambient and for

H

2

O span

CONFIG

PUMP_P_SETPT

kPa

setpoint pressure for the pump speed control

CONFIG

LEVELS_USED

number of profile levels to be used

CONFIG

MEASURE_TANKS

set = to True to automatically measure zero and span

cylinders

CONFIG

AUTO_ZEROSPAN

set = to True to zero and span the IRGA when cylinders

are measured

CONFIG

CAL_INTERVAL

min

time between automatic zero/span sequences

CONFIG

CAL_TIMEOFFSET

min

time offset for starting automatic zero/span sequences

CONFIG

CO2_SPAN_PPM

ppm

CO

2

concentration in the CO

2

span cylinder

CONFIG

BATT_LOWLIMIT

V

AP200 will shut down if supply voltage falls below this

limit

CONFIG

BATT_DEADBAND

V

AP200 will power up if supply voltage rises this much

above lower limit

C-3