Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual
Page 31
![background image](/manuals/415845/31/background.png)
AP200 CO
2
/H
2
O Atmospheric Profile System
Average power varies from 14 W above 30°C to 24 W at –30°C as
shown in FIGURE 4-23. This includes 2.0 W for the intake heaters (8
intake assemblies at 0.25 W each).
FIGURE 4-23. Average power consumption relative to temperature
Pump Module
Pump type:
Dual-head diaphragm pump with a
brushless DC motor
Mounting:
Mounted in an insulated, temperature-
controlled box inside system enclosure
Control:
Pumping speed is automatically controlled
to maintain the pump inlet pressure at the
set point
Maximum pumping speed: 9.0 liters per minute (LPM)
Pressure sensor range:
15.0 kPa to 115.0 kPa
Heater:
8.0 W, turns on/off at 2°C
Warm-up time:
Approximately 50 minutes from –30°C to
2°C
Fan:
0.7 W (turns on at 50°C and off at 45°C)
Valve Manifold
Mounting:
Mounted inside system enclosure
Inlets:
Eight air sample inlets plus one inlet for
zero, one inlet for CO
2
span, and one inlet
for H
2
O span
Connections:
0.25-in Swagelok
®
Mass Flow Sensor:
0 to 1.0 standard liters per minute (SLPM)
Heater:
8.0 W, turns on/off at 5°C
Warm-up time:
Approximately 20 minutes from –30°C to
4°C
Fan:
0.7 W; turns on at 45°C and off at 43°C
19