Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 31

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

Average power varies from 14 W above 30°C to 24 W at –30°C as

shown in FIGURE 4-23. This includes 2.0 W for the intake heaters (8

intake assemblies at 0.25 W each).

FIGURE 4-23. Average power consumption relative to temperature

Pump Module

Pump type:

Dual-head diaphragm pump with a

brushless DC motor

Mounting:

Mounted in an insulated, temperature-

controlled box inside system enclosure

Control:

Pumping speed is automatically controlled

to maintain the pump inlet pressure at the

set point

Maximum pumping speed: 9.0 liters per minute (LPM)

Pressure sensor range:

15.0 kPa to 115.0 kPa

Heater:

8.0 W, turns on/off at 2°C

Warm-up time:

Approximately 50 minutes from –30°C to

2°C

Fan:

0.7 W (turns on at 50°C and off at 45°C)

Valve Manifold

Mounting:

Mounted inside system enclosure

Inlets:

Eight air sample inlets plus one inlet for

zero, one inlet for CO

2

span, and one inlet

for H

2

O span

Connections:

0.25-in Swagelok

®

Mass Flow Sensor:

0 to 1.0 standard liters per minute (SLPM)

Heater:

8.0 W, turns on/off at 5°C

Warm-up time:

Approximately 20 minutes from –30°C to

4°C

Fan:

0.7 W; turns on at 45°C and off at 43°C