2 compile switches, 5 starting and stopping the sequence, Compile switches – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 49: Starting and stopping the sequence

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

BATT_LOWLIMIT as appropriate to protect his battery. If AC main power

and an AC/DC adapter are used, the limit may be left at 9.7 V.

BATT_DEADBAND: This variable, along with BATT_LOWLIMIT,

determine when the AP200 will restart after an automatic power shutdown.

The AP200 will not restart until the supply voltage

batt_volt reaches at least

BATT_LOWLIMIT + BATT_DEADBAND. The purpose of the deadband

(the gap between the shut-down voltage and the turn-on voltage) is to protect

the AP200 from repeated power cycles when the battery voltage is very near

the limit.

BATT_DEADBAND must be zero to 10.0 V. The default is 1.0 V.

5.4.2 Compile Switches

The AP200 program defines three constants that are used as compile switches.

These constants define which optional values are to be stored in the output

tables.

MaxLevels: MaxLevels defines the maximum number of profile levels. The

default is eight, but it may be set to the actual number of levels used (four to

eight) to avoid null data for unused levels in the

IntAvg output table. The

number of levels measured is determined by system configuration variable

LEVELS_USED.

SaveAll_diagnostics: If constant SaveAll_diagnostics is set to False (the

default), the

RawData output table will contain only those values required for

normal operation. If it is set to

True, the AP200 will save some additional

diagnostic values in output table

RawData.

N_AirTemps: Constant N_AirTemps determines the number of 107-L

temperature sensors to measure. It may be from one to eight, or zero, which is

used to disable air-temperature measurements. The default is zero. See

Appendix E, Temperature Profile, for more details on adding a temperature

profile measurement.

5.5 Starting and Stopping the Sequence

In normal operation, the AP200 will automatically cycle through the profile

inlets. If configured for automatic zero/span, it will periodically run the

calibration (zero/span) sequence and then return to the profile sequence. The

user may stop the sequence to control the valves manually and then restart the

sequence using the public variables described below. If any of the variables

that control the valve sequence are edited, the sequence must be stopped and

restarted before changes will be in effect. The following public variables allow

the user to start and stop the sequence and to control the valves manually.

seq_ACTIVE: Reports whether the AP200 is controlling the valves (True) or

not (False). If

seq_ACTIVE = False, the valves can be controlled manually.

This variable is for information only; its value cannot be directly changed by

the user.

STARTsequence: Allows the user to start the sequence. Set this variable to

True to start the automatic valve switching sequence. The AP200 will change

valve_number to 1 and wait until it can synchronize the valve sequence to the

real-time clock. This may take up to two minutes (see Appendix F, Valve

37

Advertising