Appendix F. Valve Sequence Timing

Select inlet 1 to allow at least 10 s extra equilibration time. This

allows the humidity in the tubing and IRGA to equilibrate before

restarting the profile sequence.

As shown in TABLE F-4, this minimum timing adds up to 170 s for the

zero/span sequence. The timing parameters are adjusted as needed to equal a

multiple of the profile sequence cycle time. For example, if there are 8 levels

used, the profile cycle time is 120 s. Because this is less than the 170 s

required, it is doubled to 240 s. Extra equilibration time is added to each step to

make the zero/span sequence equal 240 s. If there are four levels used, the

profile cycle time is only 60 s. Three profile cycles times are used to give 180 s

for the zero/span sequence.

TABLE F-4. Timing for Zero/Span Sequence (setting the IRGA)

Number

of Levels

Profile

Cycle

Time (s)

Measure

CO

2

Span

Time (s)

Measure

& Set Zero

Time (s)

Measure

& Set CO

2

Span

Time (s)

Equilibration

Time (s)

Zero/Span

Time (s)

Minimum

30

70

60

10

170

8

120

40

100

70

30

240

7

120

40

100

70

30

240

6

90

30

70

60

20

180

5

90

30

70

60

20

180

4

60

30

70

60

20

180

Each time the zero/span sequence is run, five records will be added to the

SiteAvg and CalAvg tables:

CO

2

Span tank before setting zero or span

Zero tank before setting zero or span

Zero tank after setting zero

CO

2

Span tank after setting zero

CO

2

Span tank after setting span