5.3.2 Intake Heater Cables

The AP200 intake assemblies have a small heater in the rain diverter to prevent

condensation. Power for these heaters is available in the AP200 system

enclosure. Campbell Scientific pn 9922 cable or equivalent is recommended

for connecting the intake assemblies to the system enclosure. The system is

designed to daisy-chain the heater cables from one intake assembly to the next

as described below and shown in FIGURE 5-16.

Make sure power is disconnected from the AP200 before

connecting the heater cable.

To bring cables into the AP200 enclosure, remove the cap from the cable

feedthrough by loosening the thumbscrew and twisting the cap while pulling it

off. See FIGURE 5-11.

FIGURE 5-11. Cable feedthrough cap shown removed to admit cables

into the system enclosure

Insert the heater cable through the feedthrough, and connect it to the DIN rail

bus in the system enclosure. The red, positive wire connects to one of the

terminals labeled

Intake Heater +. Connect the black, negative wire to one of

the terminals labeled

G, as shown in FIGURE 5-12. Two connections are

provided for convenience if multiple cables are to be connected (for example, if

the intake assemblies are mounted on two separate towers).