2 ap200 enclosure, 3 intake assemblies, Ap200 enclosure – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 33: Intake assemblies, Enclosure on ut30 tower

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

Large pole (4.0-in to 10.0-in diameter)

No mounting bracket

Consult the ENC10/12, ENC12/14, ENC14/16, ENC16/18 Instruction Manual,

available at

www.campbellsci.com

, for details on mounting bracket options.

5.1.2 AP200 Enclosure

Mount the AP200 system enclosure where it can be accessed easily to retrieve

data from the CF cards on the datalogger. The AP200 system enclosure is

similar to the ENC16/18 enclosure, and it has the same mounting options

available. Consult the ENC10/12, ENC12/14, ENC14/16, ENC16/18

Instruction Manual, available at

www.campbellsci.com

, for mounting details.

Typical installation of an AP200 system enclosure on a UT30 tower is shown

in FIGURE 5-1.

FIGURE 5-1. Installation (showing mounting hardware) of AP200

system enclosure on UT30 tower

Open the sealed bag containing the desiccant packs and humidity card. Place

two of the desiccant packs and the humidity indicator card in the mesh pocket

in the enclosure door to desiccate the inside of the enclosure. Reseal the

remaining two desiccant packs in the bag to use later.

5.1.3 Intake Assemblies

The intake assemblies are mounted at the desired air sampling positions. They

can be mounted on vertical pipes of 1.3-cm (0.5-in) to 5.1-cm (2.0-in)

diameter. The orientation of the intake assembly mounting clip must be

reversed to accommodate this range of diameters. For mounting the intake

21