AP200 CO

2

/H

2

O Atmospheric Profile System

Insert and connect a second length of heater cable for the next intake assembly.

Continue this “daisy chain” from one intake assembly to the next until all of

the intake assemblies are connected as shown in FIGURE 5-16. Intake

assemblies will accommodate a third heater power cable that may be used to

branch off to another intake assembly. This may be useful if the intake

assemblies are mounted on two or more towers.

FIGURE 5-16. Three intake assemblies with heater cables daisy

chained to the AP200 system enclosure

To check the intake heater wiring, temporarily disconnect the heater cable from

the AP200 system enclosure and measure the resistance between the red and

black wires. A hand-held digital multimeter works well for this. Each heater

has a resistance of approximately 560 ohms and should be wired in parallel.

The nominal resistance will be 560 divided by the number of heaters. TABLE

5-1 shows the nominal equivalent resistance for 4 through 8 intake heater wired

in parallel. It also gives the acceptable range, which accounts for the 5%

tolerance on the resistance of each heater. If the resistance is outside this range

it is likely there is a problem with the connections or with a heater.

The resistance of the cable is small enough that it can generally

be ignored.

NOTE

31