AP200 CO

2

/H

2

O Atmospheric Profile System

FIGURE 5-5. Tubing connections on bottom of AP200 enclosure

If the AP200 system enclosure is mounted low to the ground, it

may be difficult to read the label on the bottom of the enclosure.

For convenience, the inlet connections are also labeled on the

inside of the enclosure as shown in FIGURE 5-6.

FIGURE 5-6. Labeled inlet connections inside enclosure

Connect the lowest intake assembly to inlet

1 on the AP200 system enclosure,

the next intake assembly to inlet

2, and so forth. Connecting the tubes in this

order (shortest to longest) will minimize the time skew between intake

assemblies. If using fewer than eight intake assemblies, they must be

connected to the AP200 system enclosure inlets in order, starting with

1. Plug

unused inlets using Swagelok

®

plugs (pn 15891) that are included with the

AP200.

NOTE

24