Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 98

Advertising
background image

Appendix F. Valve Sequence Timing

time). For example, if there are eight levels used, the profile cycle time is 120

s. This means 40 s must be added to the zero/span sequence. This has been

allocated as 10 s extra (each) for the span and zero, and 20 s extra for

equilibration on level 1 at the end of the sequence.

The zero/span sequence takes the place of one profile cycle for five, six, seven,

or eight levels. However, if there are four levels used, the profile cycle time is

only 60 s. This is less than the 80 s required for a zero/span check, so two

profile cycles must be replaced. This gives a total of 120 s, the same as the

scenario with eight levels.

TABLE F-3. Timing for Zero/Span Sequence (check only)

Number

of Levels

Profile Cycle

Time (s)

Span

Time (s)

Zero

Time (s)

Equilibration

Time (s)

Total Zero/Span

Time (s)

Minimum

30

40

10

80

4

60

40

50

30

120

5

90

30

40

20

90

6

90

30

40

20

90

7

120

40

50

30

120

8

120

40

50

30

120

Similar to the profile sequence, the zero/span sequence timing parameters are

implemented as a set of three arrays:

CalSequence(), CalOnCounts(), and

CalOmitCounts(). These arrays are not accessible in the Public table, but they

are output to the TimeInfo table as a record of the system configuration. See

Appendix D, Output Tables, for details on this table.

Each zero/span sequence adds two records to the

SiteAvg and CalAvg output

tables: one for CO

2

Span and one for Zero.

With Zero and Span

If the AP200 is configured with

MEASURE_TANKS = True and

AUTO_ZEROSPAN = True, it will check the zero and CO

2

span, command

the IRGA to set them, and then verify the new readings. This section gives

timing details for this case. The AP200 zero/span sequence will:

Select inlet 10 (CO

2

Span), omit at least 20 s for equilibration, and

include 10 s in the average

Select inlet 9 (Zero), omit at least 30 s for equilibration, and include

10 s in the average

Stay on inlet 9 (Zero) and set the CO

2

zero (~10 s)

Stay on inlet 9 (Zero) and set the H

2

O zero (~10 s)

Stay on inlet 9 (Zero) and average for ~10 s

Select inlet 10 (CO

2

Span), equilibrate for 30 s, and average 10 s

Stay on inlet 10 (CO

2

Span) and set the CO

2

span (~10 s)

Stay on inlet 10 (CO

2

Span) and average for ~10 s

F-4

Advertising