Appendix F. Valve Sequence Timing

The profile sequence timing parameters are implemented in the AP200

CRBasic program as a set of three arrays. These arrays have eight locations to

accommodate up to eight levels. These arrays are not accessible in the Public

table, but they are output to the

TimeInfo table as a record of the system

configuration. See Appendix D, Output Tables, for details on this table. As

the AP200 program steps through the profile sequence, variable

sequence_index cycles from 1 to LEVELS_USED to access the values stored

in the corresponding location in these arrays.

Array ProfileSequence() contains the valve number for each step in the

profile sequence. The values in this array increment from 1 to

LEVELS_USED. If the intake assemblies are installed according to normal

convention, the lowest intake will be designated level 1 and will connect to

inlet 1 on the AP200 system enclosure. This gives a one-to-one correspondence

between public variables

valve_number and sequence_index, as well as the

terms level and step. (The intake assembly at the bottom of the tower (level 1)

connects to inlet 1 on the AP200 system enclosure, which is controlled by

valve number 1. This valve is selected during step one of the profile sequence,

defined by

seq_index = 1). Array locations beyond LEVELS_USED are

filled with zero.

Array ProfileOnCounts() contains the time (number of scans) for each step in

the profile sequence. This is similar to the

Time on Level given in TABLE

F-1, except it is multiplied by two to account for the datalogger scan rate (2

scans/s). For most cases, the same value is used for each step in the sequence.

However, for seven levels, one extra second is allocated for level 1 to achieve

an integer number of cycles per half hour. Array locations beyond

LEVELS_USED are filled with zero.

ProfileOmitCounts() is the time (number of scans) to wait after a valve switch

before including data in the average. This is the

Omitted Time given in

TABLE F-1, multiplied by two to account for the datalogger scan rate (2

scans/s). For most cases, the same value is used for each step in the sequence.

For seven levels, an extra second is allocated for level 1 to achieve an integer

number of cycles per half hour. Array locations beyond

LEVELS_USED are

filled with zero.

Synchronization to the Real-time Clock

When the AP200 program starts or when the user restarts the sequence (see

Section 5.5, Stopping and Starting the Sequence), the AP200 will delay the

actual profile sequence start by up to two minutes to synchronize the sequence

to the datalogger real-time clock. The following examples illustrate this

synchronization:

Example 1:

LEVELS_USED = 8, STARTsequence set to True at 9:15:25:

As shown in TABLE F-1, the cycle time for 8 levels is 120 s (2 min). To

synchronize with the real time clock, the sequence must start at a multiple of 2

min, (for example, 9:00:00, 9:02:00, 9:04:00, etc.) Valve 1 will be selected as

soon as

STARTsequence is set = True (9:15:25), but the sequence will not

actually start until 9:16:00 (the earliest multiple of 2 min). Data will be

included in the average for level 1 from 9:16:10 to 9:16:15, when the valve will

switch to level 2. Data will be included in the average for level 2 from 9:16:25

to 9:16:30, when the valve will switch to level 3, and so on.