2 enclosure desiccant, 3 intake filters, Enclosure desiccant – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 59: Intake filters

Advertising
background image

AP200 CO

2

/H

2

O Atmospheric Profile System

Some common causes of leaks include:

Forgetting to tighten a fitting during the installation. It is good

practice to check all of the tube connections before turning the system

on. If a fitting can be loosened with fingers, it is not properly

tightened.

Dirt or debris in the tube fitting. Use care when connecting or

disconnecting tubes, and install caps and plugs on tubes and fittings

when they are not in use.

Excessive bending. Be careful to observe the minimum bend radius

and avoid bending the tube near the end at the connections.

7.2 Enclosure Desiccant

Check the humidity indicator card in the mesh pocket in the AP200 system

enclosure door. The humidity indicator card has three colored circles that

indicate the percentage of humidity. Desiccant packets inside the enclosure

should be replaced with fresh packets when the upper dot on the indicator

begins to turn pink. The indicator card does not need to be replaced unless the

colored circles overrun.

Campbell Scientific strongly suggests replacing desiccant

instead of reactivating old desiccant. Improper reactivation

can cause the desiccant packets to explode. If the user is

determined to reactivate old desiccant packets, follow the

procedure provided in the ENC16/18 User Manual.

The replacement desiccant pack is the 4905 Single 4-Unit Silica Desiccant

Bag. (The 4905 can be purchased in quantities of 20 as pn 6714.) The

replacement humidity indicator card is pn 28878.

If the desiccant packs are being replaced very frequently, make sure the

feedthrough cap is properly installed (see Section 5.3.3, Power, and FIGURE

5-19). In very humid conditions, it may be helpful to seal the cable feedthrough

with plumber’s putty.

7.3 Intake Filters

The sample flow will decrease over time as particulates clog the intake filters.

Eventually, the flow will be reduced to the extent that it will degrade the

equilibration time after an intake is selected. As a general guideline, the filters

should be replaced when the flow decreases by 25%. The filters will normally

last a few months, but will require more frequent changes in conditions with

high particulate matter in the air.

To change the intake filter, first turn the pump off to avoid pulling dirty air into

the system.

Open the cover of the intake assembly as shown in FIGURE 5-14. The rain

diverter, intake assembly with filter, and intake tubing are shown in FIGURE

5-15.

Press the rain diverter upwards and out of the bracket as shown in FIGURE

7-1.

CAUTION

47

Advertising