Appendix C. Public Variables

TABLE C-1. Public Variables

Usage

Variable Name

Units

Description

INFO

seq_ACTIVE

Boolean flag: True if the valve sequence is active

INFO/USER valve_number

valve number (1 to 11); determines which inlet is selected

INFO

sequence_index

index for valve switching sequence; cycles from 1 to

LEVELS_USED

INFO

smpl_counter

number of scans since last valve switch

INFO

end_omit

Boolean flag: True when equilibration time after valve

switch is satisfied

DIAG

panel_tmpr

°C

temperature of the datalogger wiring panel

DIAG

batt_volt

V

supply voltage; measured inside datalogger after diode

voltage drop

DIAG

batt_volt_LOW

Boolean flag: True if batt_volt is too low (controls power

shutdown)

DIAG

valve_tmpr

°C

temperature of the valve manifold

DIAG

valve_tmpr_OK

Boolean flag: True if the valve module is within operating

limits

DIAG

valve_heat_ON

Boolean flag: True if the valve heater is on

DIAG

valve_fan_ON

Boolean flag: True if the valve fan is on

USER

pump_ON

set = to True to enable the sample pump

DIAG

pump_press

kPa

pressure measured at the pump inlet

DIAG

pump_control

pump speed control variable; can be from 0 (off) to 1 (full

speed)

DIAG

pump_speed

Hz

pump speed; measured by a tachometer on the pump

DIAG

pump_tmpr

°C

temperature of the pump

DIAG

pump_tmpr_OK

Boolean flag: True if the pump temperature is within its

operating range

DIAG

pump_heat_ON

Boolean flag: True if the pump heater is on

DIAG

pump_fan_ON

Boolean flag: True if the pump fan is on

INFO

intake_heat_ON

Boolean flag: True if the intake heaters are on

TAIR

T_air(1)

°C

air temperature profile (optional, defined only if

N_AirTemps > 0)

TAIR

T_air(2)

°C

air temperature profile (optional, defined only if

N_AirTemps > 1)

TAIR

T_air(3)

°C

air temperature profile (optional, defined only if

N_AirTemps > 2)

TAIR

T_air(4)

°C

air temperature profile (optional, defined only if

N_AirTemps > 3)