Appendix B. AP200 Diagnostics

The public variable

diag_AP200 indicates conditions outside the normal

operating range for the AP200. Any value other than zero indicates a problem.

Diag_AP200 contains several error flags encoded as binary bits. To diagnose a

problem, first decode the value according to TABLE B-1 and then see the

corresponding section in this appendix.

TABLE B-1. Summary of Bit Numbers Indicating Conditions Outside Normal Operating Range

Binary Bit (LSB = 1)

Numeric Value

Error Condition

1

1

Battery voltage is too low

2

2

Valve temperature is outside its operating range

3

4

Pump temperature is outside its operating range

4

8

Pump pressure is not at the setpoint

5

16

Sample flow is outside its normal range

6

32

Cell temperature is outside its operating range

7

64

Cell pressure is not at the setpoint

Example 1:

Assume

diag_AP200 = 34

Because 34 = 32 + 2, this indicates bit 6 (numeric value = 32) and bit 2

(numeric value = 2) are set. This means there are two problems:

Bit 6: Cell temperature is outside its operating range

Bit 2: Valve temperature is outside its operating range

This could arise if the system is powered up in cold weather, and the valve

manifold and IRGA sample cell have not yet reached their operating

temperatures.

Example 2:

Assume

diag_AP200 = 16

This indicates bit 5 (numeric value = 16) is set. This means the sample flow is

outside its normal range of 100 to 300 ml/min.

Output Tables

Every sample of

diag_AP200 is stored in the RawData output table.

However, only an average of

diag_AP200 is stored in the output tables

IntAvg, CalAvg, and SiteAvg. The interpretation of this average, given as

diag_AP200_Avg, is slightly different than the original variable, diag_AP200,

for two reasons. First, it represents only the samples that are included in the

averages (see Appendix F, Valve Sequence Timing, for timing details).