FIGURE 5-7 shows the tubing connections from the intake assemblies to the

system enclosure.

FIGURE 5-7. Tubing connections from four intake assemblies

connected to inlets 1 – 4

5.2.2 Zero and CO

2

Span

The AP200 can perform automated zero (CO

2

and H

2

O) and CO

2

span of the

IRGA. This requires the user to supply cylinders of zero air and CO

2

span gas,

with appropriate regulators.

Use high-quality gases for the zero and CO

2

span. The zero gas

must be free of significant water vapor and CO

2

. The CO

2

span

gas should have a well-known concentration of CO

2

balanced in

air (not nitrogen).

Install these cylinders in close proximity to the AP200 system enclosure. Each

cylinder must have a pressure regulator to control the outlet pressure at 0 psig

and must have a 0.25-in Swagelok

®

fitting on the outlet. Connect these fittings

to the valve module inlets using 0.25-in OD tubing, such as pn 15702.

Minimize the length of these tubes to reduce the equilibration time after the

zero or CO

2

span cylinder is selected. FIGURE 5-8 illustrates this

configuration. Refer to Appendix H, Using Swagelok

®

Fittings, for

information on installing and replacing Swagelok

®

fittings. For convenience,

Campbell Scientific can supply pre-swaged tube assemblies (pn 21823-L) for

this purpose.