Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 75

Advertising
background image

Appendix B. AP200 Diagnostics

If the pump pressure is too low, check

pump_control and pump_speed.

When the pressure falls below the setpoint, the AP200 will respond by

reducing

pump_control. This should reduce pump_speed and allow

pump_press to rise to the setpoint. If pump_control = 0 but the pump is still

running, contact Campbell Scientific.

If the pump speed is oscillating, check the value of

BuffDepth. This variable

is found in the Status table (see the CR1000 User Manual at

www.campbellsci.com

for more information). The AP200 CRBasic program

runs in Pipeline mode, which allows measurements to stay on schedule even

when processing tasks fall behind.

BuffDepth indicates how far processing

has fallen behind measurements (number of scans). If processing falls too far

behind, the delay between measuring the pump pressure and adjusting the

pump speed can cause the pump speed to oscillate. Normally,

BuffDepth will

be 2 or less. If the value is higher than 2, contact Campbell Scientific.

Bit 5

: Sample flow is outside its normal range

If bit 5 of

diag_AP200 is set, this indicates the sample flow is outside the

normal range. To confirm this error, check the value of

sample_flow which is

measured by a mass-flow sensor mounted on the outlet of the valve manifold.

The sensor measures the flow to the IRGA. The normal range for the sample

flow is 100 to 350 ml/min. If the sample flow is outside this range, see the

appropriate section below.

Profile Inlets

The sample flow for profile inlets is set by an orifice in the intake assembly.

See Section 4.2.1, Intake Assemblies, for details.

If the flow goes too high at a valve switch but comes into the normal range

within a few seconds, this is normal. It is caused by a different pressure drop

through the valve manifold depending on whether the inlet is selected or

bypassed. The combined flow from the bypassed inlets causes slightly more

pressure drop than for the inlet selected to go to the IRGA. Consequently, the

pressure in the intake tubes and mixing volumes increases slightly when they

are bypassed. When a bypassed inlet is selected, the pressure equalizes by

pushing slightly higher flow than normal through the IRGA. This is acceptable

as long as the sample flow is within the normal range by the time the data are

included in the average.

If the flow for one or more of the inlets stabilizes at a high flow, this may

indicate a leak in a tube or connection. See Section 7.1, Plumbing Leaks, for

leak-checking suggestions.

If the flow is too low for one or more inlets, this is most likely caused by a

filter that is plugged with particulates. This normally happens over a span of

weeks or months. It may happen more quickly in dirty conditions. Try

replacing the filter (see Section 4.1.6, Replacement Parts). If this does not

resolve the problem, the orifice may be plugged, and Campbell Scientific

should be contacted.

B-5

Advertising