3 sequence of a determination, Sequence of a determination, Figure 3 – Metrohm 920 Absorber Module User Manual
Page 17: Sequence of the determination, Figure 3 sequence of the determination
Advertising
■■■■■■■■■■■■■■■■■■■■■■
2 The Combustion IC system
920 Absorber Module
■■■■■■■■
9
2.3
Sequence of a determination
A determination in the Combustion IC system runs according to the fol-
lowing diagram:
Start determination
(CIC system)
Rinse absorber tube and
provide absorber solution
(920 Absorber Module)
Sample input
(Combustion Module)
Rinse capillary f. combustion
gases w. absorber solution
(920 Absorber Module)
Absorption of
combustion gases
(920 Absorber Module)
Optional: Water infeed
(920 Absorber Module)
Combustion
(Combustion Module)
Fill loop with
absorption solution
(920 Absorber Module)
Matrix elimination
(930 Compact IC Flex)
Re-rinse absorber tube
(920 Absorber Module)
IC determination
(930 Compact IC Flex)
Figure 3
Sequence of the determination