Figure 20, Port valve – valve positions – Metrohm 920 Absorber Module User Manual
Page 74
4.6 6-port valve
■■■■■■■■■■■■■■■■■■■■■■
66
■■■■■■■■
920 Absorber Module
3
Port 3
For connecting the sample loop.
4
Port 4
For connecting the capillary to the ion chro-
matograph.
5
Port 5
For connecting the capillary from the outlet
of the Metrosep I trap column.
6
Port 6
For connecting the sample loop.
Valve positions
The 6-port valve can have two valve position:
1
2
Figure 20
6-port valve – Valve positions
1
Filling
2
Injecting
■
Filling
In the Filling position, ports 2 and 3, 4 and 5, and 6 and 1 are con-
nected to each other.
The sample flows through port 1 into the 6-port valve, through the
sample loop and through port 2 out of the 6-port valve into the waste
container.
At the same time, the transfer solution flows through port 5 into the 6-
port valve and directly back out through port 4.
The sample loop is filled with the sample at this valve position.