Figure 18, Connecting the 6-port valve – Metrohm 920 Absorber Module User Manual

Page 72

Advertising
background image

4.6 6-port valve

■■■■■■■■■■■■■■■■■■■■■■

64

■■■■■■■■

920 Absorber Module

Connecting the 6-port valve

6.1803.030

6.1803.160

6.1803.030

6.2744.010

6.2744.010

6.2744.010

6.2744.010

6.1831.040

6.1825.290

1

2

3

4

5

6

Figure 18

Connecting the 6-port valve

1 Connecting port 1

Tighten the loose end of the PEEK capillary connected to port 1 of
the 10-port valve (17-2) to port 1 of the 6-port valve using a pres-
sure screw.

2 Connecting port 2

Push one end of the PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160), (17-1) through a capillary feed-through below the
bottle holder from the back towards the front of the instrument.

Push the pressure screw over the end of the PTFE capillary far
enough that approx. 4 mm of the capillary is exposed.

Screw the pressure screw in place at port 2 of the 6-port valve.

Shorten the other end of the capillary using a capillary cutter so
that it reaches the waste container.

Fasten the end to the waste container.

3 Connecting port 4

Push one end of the PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030), (17-18) through the cable feed-through at the bot-
tom left on the front of the instrument and then route it up to the
6-port valve through the two capillary guides on the left inside the
instrument.