4 Installation

920 Absorber Module

■■■■■■■■

41

CAUTION

Please read through the correct procedure in the manual for the 800
Dosino before you attach the Dosino on the dosing unit.

Plugging in the connection cable

1 Route the connection cable for the Dosino for the absorber solution

out of the instrument through the cable opening (6-1) in the back
panel.

2 Plug the plug for the connection cable into the MSB 1 socket.

4.4.2

Connecting the ports – Overview

Using ports

The ports on the Dosino for the absorber solution are used as follows:

Port …

… is used for …

… and is connected to

1

Liquid Handling

T connection

2

Aspirating the absorber solution

Bottle with absorber solution

3

Dosing the absorber solution

Metrosep A Trap 1 - 100/4.0
(6.1014.000) trap column inlet

4

Not used

Accessories

The following accessories have to be connected to the Dosino for the
absorber solution:

Port

Install this accessory

Port 1

FEP tubing M6 for 920, 9 cm (6.1805.040) (10-

12

)

Port 2

FEP tubing M6 for 920, 150 cm (6.1838.020) (10-

11

)

Port 3

PTFE capillary (6.1803.030), approx. 20 cm (10-

10

)

Pressure screw (6.2744.350)

Coupling (6.2744.080)

Port 4

For this step you also need:

Thread adapter (6.1618.020)

Dosino holder (6.2057.230)

Capillary cutter (6.2621.080)