Figure 23, Combustion tube – outlet – Metrohm 920 Absorber Module User Manual

Page 84

Advertising
background image

4.10 Connecting the 920 Absorber Module to the remaining components

■■■■■■■■■■■■■■■■■■■■■■

76

■■■■■■■■

920 Absorber Module

4.10.2

Connecting the Combustion Module outlet: Absorber solution

The capillary with the yellow colored sleeve and the thicker capillary are
connected to the CIC T connector in the Combustion Module.

There are 2 connections on the CIC T connector in the Combustion Mod-
ule:

Figure 23

Combustion tube – Outlet

1

Absorber solution connection
Connection for the capillary with the yellow
colored sleeve (7-

9

).

2

T connector CIC (6.7301.210)
T connector on the combustion tube's out-
let. For feeding the absorber solution in and
removing the combustion gases.

3

Peltier cooler
For cooling the combustion gases at the
combustion tube's outlet.

4

Combustion gases connection
Connection for the PTFE capillary, 1/8 in. /
1/16 in. ID / 1.5 m (6.1803.130) (7-

8

).