Metrohm 920 Absorber Module User Manual

Page 92

Advertising
background image

6.9 Absorber tube

■■■■■■■■■■■■■■■■■■■■■■

84

■■■■■■■■

920 Absorber Module

Exchanging the absorber tube holder

1 Unscrewing the outlet capillary

On the bottom side, unscrew the pressure screw from the absorber
tube holder.

2 Unscrewing the check valve holder

Unscrew the check valve holder (6.2766.010) from the absorber tube
holder.

3 Replacing the absorber tube holder

Unscrew the absorber tube holder from the absorber tube.

Place a new sealing ring in the new absorber tube holder.

On the bottom, screw the new absorber tube holder to the
absorber tube.

4 Tightening the check valve holder

Screw the check valve holder with the new inlet valve to the side of
the absorber tube holder.

5 Fastening the inlet capillary again

Screw the inlet capillary (6.1803.130) to the check valve again using
the pressure screw (6.2744.170).

6 Mounting the outlet capillary

Tighten the loose end of the capillary (6.1803.160) (connected to
port 8 of the 10-port valve) to the bottom of the absorber tube
holder using a pressure screw (6.2744.010).

Advertising