Figure 14, Connecting the 10-port valve, Figure 14 connecting the 10-port valve – Metrohm 920 Absorber Module User Manual

Page 64: 1 connecting the central port, Screw the adapter in place in the central port, 2 connecting port 1

Advertising
background image

4.5 10-port valve

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

920 Absorber Module

Port

Install this accessory

8

PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 60 cm
(13-

5

)

Pressure screw (6.2744.090)

9

PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 5 cm
(13-

6

)

Pressure screw (6.2744.090)

10

PTFE capillary, 0.75 mm ID / 3 m (6.1803.160) (13-

7

)

Pressure screw (6.2744.090)

Connecting the 10-port valve

Connect the ports of the 10-port valve according to the following figure:

NOTE

We recommend connecting the central port first.

6.2744.090

6.1831.040

6.2744.090

6.1803.160
80 cm

6.1803.160
80 cm

6.2744.090

6.27440220

6.2744.220

6.2744.220

6.2744.220

6.2744.090

6.1803.160

60 cm

6.2744.090

6.1803.160

5 cm

6.2744.090

6.1803.160
3 m

6.2744.200

6.1462.130

1

2

3

4

5

6

7

8

9

10

Figure 14

Connecting the 10-port valve

1 Connecting the central port

Screw the adapter in place in the central port.

Screw one end of the transfer tubing in place on the adapter.

2 Connecting port 1

Screw the PEEK capillary to port 1 using a long pressure screw.