4 Installation
920 Absorber Module
■■■■■■■■
33
4.3.4
Inserting the Dosino in the instrument
CAUTION
Space is limited in the 920 Absorber Module.
When you insert the fully connected Dosino into the instrument, ensure
that none of the tubing or capillaries are bent or kinked. Turn the Dos-
ino so that the capillaries and tubing can be as straight as possible.
Inserting the Dosino in the instrument
1.
2.
1 Place the Dosino holder onto the mounting bolts and fix it in place
using the knurled screw.
Next step
Connect
capillary/tub-
ing at port
…
… to
1
Metrosep I Trap 1 - 100/4.0 (6.1014.200) trap column inlet (see
Chapter 4.3.5, page 34)
2
Water bottle (see Chapter 4.3.6, page 36)