Installing the drainage tubing
Liquid that leaks in the bottle holder flows to the rear of the instrument
and is drained through an opening on the bottle holder. Drainage tubing
has to be mounted at this opening. This drainage tubing guides the leak-
ing liquid to the base tray where the leak sensor is located.
Accessories
For this step you need the following parts from the accessory kit: Vario/
Flex Basic (6.5000.000):
■
2 × silicone tubing (6.1816.080)
You also need scissors.
Connecting the drainage tubing
6.1816.080
6.1816.080
1 Attach one end of the silicone tubing to the drainage support on the
bottle holder.
Attach the other end of the silicone tubing to the right-side drainage
support on the base tray.
2 Attach one end of the second piece of silicone tubing to the left-side
drainage tubing connection on the base tray.