Metrohm 920 Absorber Module User Manual

Page 34

Advertising
background image

4.3 Dosino for ultrapure water

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

920 Absorber Module

Connecting port 2 and port 4
Port 2 and port 4 of the Dosino are on the bottom of the instrument.
These ports are connected first.

Accessories

The following accessories have to be connected to these two ports:

Port

Install this accessory

2

FEP tubing M6 for 920, 150 cm (6.1838.020) (8-

15

)

4

PTFE capillary (6.1803.030) (8-

14

)

Pressure screw (6.2744.350)

Coupling (6.2744.080)

Adapter Dosino port 4 (6.1808.280)

For this step you also need:

Thread adapter (6.1618.020)

Dosino holder (6.2057.230)

Colored sleeves, blue

Preparing

Before you can connect the capillary and the aspiration tubing, you have
to:

Install the coupling for the connection between the capillary and Dos-
ino.

Thread the capillary for the water infeed and the aspiration tubing
through the thread adapter and the Dosino holder.

1 Mounting the coupling

The ports on the Dosino are intended for tubing with an M6 connec-
tor. If capillaries are connected to the ports, a coupling has to be
installed at the end.

6.2744.080

6.2744.350

6.1803.030

Optional: Guide a blue colored sleeve over the end of the capil-
lary.