3 connecting the bottle with the check standard, Connecting the bottle with the check standard – Metrohm 920 Absorber Module User Manual

Page 67

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

920 Absorber Module

■■■■■■■■

59

Inserting the bottle with the standard solution into the
instrument

1 Insert the bottle with the standard solution into the bottle holder.

4.5.3

Connecting the bottle with the check standard

The bottle with the standard solution is connected to the 10-port valve.

Connecting the bottle with the check standard

1 Guide the loose end of the capillary (13-4) (connected to port 3 of

the 10-port valve) in through a capillary feed-through between the
device and the bottle holder on the side of the instrument.

2 Fasten the capillary in the bottle containing the check standard.

Place the bottle on the bottle holder.