4 Installation
920 Absorber Module
■■■■■■■■
43
Installing the water aspiration tubing at port 2
1 Threading the Dosino holder thread adapter
Before you can connect the aspiration tubing for the absorber solu-
tion, you have to thread the aspiration tubing for the absorber solu-
tion through thread adapter and the Dosino holder.
6.1838.020
6.2057.230
6.1618.020
■
Start by threading the FEP tubing through the thread adapter
from below and then through the Dosino holder.
2 Connecting the FEP tubing to port 2
Tighten the FEP tubing on port 2 of the dosing unit.
3 Routing the FEP tubing out of the instrument
Guide the loose end of the FEP tubing out of the 920 Absorber Mod-
ule through one of the capillary openings on the back panel of the
instrument.
Tightening the Dosino holder on the Dosino for the
absorber solution
Prerequisites: