Figure 12 – Metrohm 920 Absorber Module User Manual

Page 58

Advertising
background image

4.4 Dosino for the absorber solution

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

920 Absorber Module

Feed the capillary for the absorber solution into the instru-
ment

The capillary for the absorber solution is fed into the instrument as fol-
lows:

6.1803.030

Figure 12

Feed the capillary for the absorber solution into the instru-
ment

1

Slide a yellow colored sleeve over the capillary far enough that it
remains visible outside the instrument.

Route one end of the PTFE- capillary (6.1803.030) to the front of
the instrument from the right side of the instrument through the
capillary feed-through between the base tray and the 920
Absorber Module.

Thread the end of the capillary through the bottom horizontal
capillary guide on the right inner side wall and then through the
vertical capillary guide from the bottom to the top.

Optional: Fasten another yellow colored sleeve to the end of the
capillary.

Connecting the Metrosep A Trap 1 – 100/4.0

NOTE

The trap column is sealed with threaded stoppers at both ends.

Remove these threaded stoppers before connecting the capillaries.

Advertising