6 Notes – Maintenance – Faults

761 SD Compact IC / Instructions for Use 8.

761.1043

98

6.2.6

Regeneration of the suppressor module

Regeneration of a suppressor operating at reduced

capacity
If the suppressor units are exposed to certain heavy metals (e.g. iron)

or organic contaminants for long periods of time, these can no longer

be completely removed by the regeneration solution (100 mmol/L

H

2

SO

4

) normally used. This diminishes the capacity of the suppressor

units which, in milder cases, results in a reduced sensitivity to phos-

phates and, in severe cases, results in a strong increase in the baseline

and/or pressure rise. If such capacity problems occur at one or several

positions, the corresponding suppressor units must be treated as fol-

lows:

1 Disconnect the suppressor from the IC system

Disconnect the suppressor from the separating column and

detector.

2 Regenerate the suppressor

Rinse each suppressor unit for approx. 15 minutes with one

of the following solutions:

Contamination with heavy metals

1 mol/L H

2

SO

4

+ 0.1 mol/L oxalic acid

Contamination with organic cationic complexing

agents

0.1 mol/L H

2

SO

4

/ 0.1 mol/L oxalic acid / acetone 5

%

Severe contamination with organic substances

0.2 mol/L H

2

SO

4

/ acetone

20 %

3 Connect the suppressor to the IC system

Reconnect the suppressor to the IC system. If capacity prob-

lems persist, replace the suppressor rotor (see Section 6.2.8).