2.4 Connection of the detector block

761 SD Compact IC / Instructions for Use 8.

761.1043

15

2.4

Connection of the detector block

The metal-free detector block 1.732.0420 which must be fitted in the

instrument and connected belongs to the scope of delivery of the 761

SD Compact IC. Proceed as follows:

1 Note the cell constant

The cell constant c = XX.X /cm, measured at the works, is

printed on the rear side of the detector block. Note this value;

it must subsequently be entered in the software in order to

ensure that an exact display of the conductivity is obtained

(see Section 2.12.3).

2 Install detector block

Unscrew the two upper knurled screws 11 from the upper

rear panel 12 of the 761 SD Compact IC, slacken the two

lower knurled screws 11 a little and remove the rear panel

(see Figure 2).

Position the detector block 40 from the rear onto the space

provided in the 761 SD Compact IC and push it fully to the

front (see Figure 7).

3 Connect detector block

Pull the grey connecting cable permanently attached to the

detector block 40 out through the opening (where the upper

rear panel 12 previously was) on the rear of the 761 SD

Compact IC and plug it in to connection 20 "Detector Block"

of the 761 SD Compact IC.

Since other connections need to be routed through the rear openings

of the 761 SD Compact IC later on, you should not fit the rear panel

12

back on until the end of the installation procedure (see Section 2.10).

Figure 7:

Position of the detector block

40

39

Inlet capillary to detector block

PEEK capillary, permanently attached

40 Detector block 1.732.0420

41

Outlet capillary from detector block

PEEK capillary, permanently attached

40

39

41