2 metrosep a supp 1 hs separating column, Metrosep a supp 1 hs separating column, Tion 2.15.2 – Metrohm 761 SD Compact IC User Manual

Page 48

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

40

2.15.2 Metrosep A Supp 1 HS separating column

Now fit the supplied Metrosep A Supp 1 HS as the actual IC separating

column.

When fitting the column, always ensure that it is fitted correctly as

shown by the flow direction on the sticker (the arrow must point in the

flow direction). Ensure that the set flow rate is not higher than the per-

mitted flow rate for the corresponding column (see the sheet en-

closed with the column).

The separating column 36 has two connections for PEEK capillaries

and must be fitted as follows:

1 Connect the separating column 36 downstream of

the precolumn 34

Detach the screw caps from the Metrosep A Supp 1 HS

separating column 36.

Screw the inlet end of the separating column 36 (note the

flow direction, the arrow on the sticker must point in the flow

direction) to the free end of the connector 35 attached to the

precolumn 34.

2 Prepare for rinsing the separating column

Position the beaker beneath the column outlet.

Open the

startup.smt

system and connect it as described in

Section 2.13.1.

Double-click on the 761 image in

the system window. You will see

the window for manual operation

of the 761 SD Compact IC:

3 Rinse the separating column

Switch on the high-pressure pump (

IC pump

) by clicking on

<On>

and rinse the column for approx. 10 min with eluent.

Switch the high-pressure pump back off again by clicking on

<Off>

.

4 Connect the separating column to the suppressor

The column outlet is connected to the eluent inlet of the sup-

pressor so as to close the eluent path:

Fit a PEEK pressure screw 50 to the "suppressor inlet capil-

lary for eluent ("Eluent")" 37.

Then screw this pressure screw into the column connection

of the separating column 36.

5 Fix the separating column in position

Insert a column holder (6.2027.040) 67 into the mounting rails

66 and secure the separating column 36 in the column hol-

der.

Advertising