3 eluents, 4 peristaltic pump, 5 suppressor module – Metrohm 761 SD Compact IC User Manual

Page 97: Eluents, Peristaltic pump, Suppressor module

Advertising
background image

6.1 Practical notes on ion chromatography

761 SD Compact IC / Instructions for Use 8.

761.1043

89

Maintenance
An unstable baseline (pulsation and flow rate fluctuations) is attributable

to contaminated valves or defective, leaking piston seals in many

cases.
Contaminated valves are cleaned by rinsing with water, RBS solution

or acetone (see Section 6.2.5). When the cleaned valves are reinstalled,

you must ensure that the flow direction is correct.
The procedure for replacement of piston seals is described in Sec-

tion 6.2.5.
Salt crystals between the piston and the seal are the cause of abrasive

particles which can contaminate the eluent. These lead to contaminated

valves, pressure rise and, in extreme cases, to scratched pistons. It is

thus essential to ensure that no precipitates can occur.

6.1.3 Eluents

The eluent for determining soft drinks is pre-prepared and packed in

MPaks. If you wish to use your own eluents, please follow the conven-

tional rules for eluent handling for IC.

6.1.4 Peristaltic

pump

The pump tubing used by the peristaltic pump is consumable material

with a limited lifetime and should be exchanged at regular intervals

(approx. every 4 weeks under continuous use; see Section 6.2.9).
The working life of pump tubing depends to a considerable extent on

contact pressure. This is why the contact pressure must be correctly set

as described in Section 2.14 and the tubing cartridges 57 must be fully

raised by releasing the snap-action lever 59 on the right-hand side if the

pump is to remain switched off for a lengthy period of time (this means

that the optimum contact pressure, once set, remains unchanged).

6.1.5 Suppressor

module

Protection
To avoid contamination of the suppressor module by foreign particles

or bacterial growth, the two filter units PEEK 56 (6.2821.120) must be

fitted upstream of the suppressor module’s inlet capillaries (see Section

2.3.3).

Operation
The Metrohm Suppressor Module MSM consists of a total of 3 sup-

pressor units which are in turn used for suppression, regenerated with

sulphuric acid and rinsed with back-flowing eluent. In order to record

every new chromatogram under comparable conditions, work is nor-

mally carried out with freshly regenerated suppressor. Switchover oc-

curs automatically together with valve switchover to

Fill

.

The suppressor units must never be regenerated (with H

2

SO

4

) in the

same flow direction used for the eluent.

Advertising