6 Notes – Maintenance – Faults
761 SD Compact IC / Instructions for Use 8.
761.1043
90
The suppressor module must
never be switched in dry state as there
is a danger of blocking.
Maintenance
In the event of reduced capacity or high counter-pressure, the suppres-
sor module must be regenerated (Section 6.2.6), cleaned (Section
6.2.7) or exchanged (Section 6.2.8).
6.1.6 Connections
All connections between injector, column and detector must be as short
as possible, have a low dead volume and must be absolutely tight.