15 precolumn and separating column, 1 metrosep rp guard, Precolumn and separating column – Metrohm 761 SD Compact IC User Manual

Page 47: Metrosep rp guard, N 2.15 )

Advertising
background image

2.15 Precolumn and separating column

761 SD Compact IC / Instructions for Use 8.

761.1043

39

2.15 Precolumn and separating column

2.15.1 Metrosep RP Guard

Using the precolumn (Metrosep RP Guard 34) with easily exchangeable
filters serves to protect the separating column 36 and considerably pro-
longs its service life.
The precolumn has two connections for PEEK capillaries and must be

fitted as follows:

1 Remove the PEEK coupling between column con-

nection capillary 33 and suppressor

Undo the PEEK coupling 54 fitted in Section 2.6.4 between

column connection capillary 33 and suppressor and remove

it. It was fitted only for rinsing without column (Section 2.14).

2 Connect the Metrosep RP Guard 34

Detach the screw caps from the Metrosep RP Guard.

Attach a PEEK pressure screw 50 to the column connection

capillary 33.

Firmly screw the precolumn to the column connection capil-

lary 33 (the arrow on the precolumn must point in the flow di-

rection).

Connect the connection capillary 35 belonging to the precol-

umn to the other end of the precolumn.

3 Prepare for rinsing the precolumn

Route the free end of connection capillary 35 connected to

the precolumn 34 into a beaker.

Open the

startup.smt

system and connect it as described in

Section 2.13.1.

Double-click on the 761 image in

the system window. You will see
the window for manual operation
of the 761 SD Compact IC (see
below):

4 Rinse the precolumn

Switch on the high-pressure pump (

IC pump

) by clicking on

<On>,

rinse the precolumn for approx. 10 minutes with eluent.

Switch the high-pressure pump back off again by clicking on

<Off>

.

Next connect the Metrosep A Supp 1 HS separating column 36, see
Section 2.15.2.