2 control with «ic cap» (manual operation), Control with «ic cap» (manual operation) – Metrohm 761 SD Compact IC User Manual

Page 89

Advertising
background image

5.1 Operation with manual sample change

761 SD Compact IC / Instructions for Use 8.

761.1043

81

2 Allow shutdown.smt to run

Insert the free end of the sample aspirating tubing 3 into a

vessel filled with water.

Start the system by selecting

SYSTEM /

Control

/ Start determi-

nation

(see Software Instructions for Use «IC Net 2.3», Section

4.3.3).

Let the system run to the end (duration: 7.10 minutes). High-
pressure pump and peristaltic pump are switched off auto-
matically at the end.

5.1.2

Control with «IC Cap» (manual operation)

Launch the «IC Cap» software and log in (see Section 2.12.4).

Configuration

manual.cfg

should have been installed during installation

(see Section 2.12.2). If not, configuration

manual.cfg

should now be

loaded (see Section 4.2.1).

5.1.2.1

Conditioning (manual operation / «IC Cap»)

The instrument should be conditioned after each restart:

1 Settings

Click on the yellow <Sample> button at the top right so that

it changes to the green <Start> button.

Choose System

startup.smt

for item

System

.

Choose type "Sample" for item

Run Type

.

Enter "Startup" for item

Ident

.

2 Condition the instrument for 30 minutes

Insert the free end of the sample aspirating tubing 3 into a

container filled with water.

Start the system by clicking on the green <Start> button

and allow it run through to the end (30 minutes). Clicking on

the <Chrom> button displays the

CHROMATOGRAM

win-

dow.

The baseline should be stable at the end.