3 description of the connections, 1 connection of capillaries/tubing, Description of the connections – Metrohm 761 SD Compact IC User Manual

Page 20: Connection of capillaries/tubing

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

12

2.3

Description of the connections

2.3.1

Connection of capillaries/tubing

The connections for eluent, sample and regenerant consist of:

PEEK capillaries 6.1831.010 (inner diameter = 0.25 mm)

PTFE microcapillaries 6.1803.030 (inner diameter = 0.5 mm)

PTFE tubing connections to the MPaks 6.1837.000 (inner di-

ameter = 1.5 mm)

The PEEK capillaries and PTFE microcapillaries can be connected ei-

ther with PEEK pressure screws 50 6.2744.010 (long) or PEEK

pressure screws 51 6.2744.070 (short, for connection to the IC

pump).

The "PTFE tubing connection to the MPak" 23 can be connected to the

PEEK pressure screw 52 4.422.4510 (wide).

The connectors must be fitted to the capillaries as follows in this case:

Capillaries provided with new connectors must feature a flawless, flat

cut edge. In order to ensure this, it is best to use the

tubing cutter

6.2621.080.

1 Fit pressure screw

Slide the corresponding pressure screw (50, 51 or 52), as

shown in Figure 4 onto the capillary 53 (23 for MPak connec-

tion).

2 Insert the capillary into the connection

Slide the end of the capillary fully into the corresponding

connector (in order to avoid dead volume).

3 Tighten the pressure screw

Firmly tighten the pressure screw (50, 51 or 52) by hand (do not

use tools).