2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

14

Capillaries may also be connected together via PEEK inline filters, see

Section 2.3.3.

2.3.3

Filter unit PEEK

Three PEEK filter units 6.2821.120 should be fitted in the 761 SD

Compact IC.
The first one should be fitted between IC pump head 24 and pulsation

dampener 30. It serves to avoid contamination by abrasive particles of

the piston seals.
The two other filter units are installed upstream of the suppressor mod-

ule. One upstream of the inlet capillary 42 (labelled "H2O") in the return

path from the detector to the suppressor in the eluent path (see Section

2.6.6). The other one upstream of the inlet capillary 45 (labelled

"H2SO4"), between peristaltic pump and suppressor module in the re-

generant path (see Section 2.7.2). They serve to protect the suppressor

module against foreign particles and bacterial growth.

For the connection of the filter unit, please note the flow direction ar-

row printed on the housing.

Figure 6:

Connecting the filter unit PEEK

50 Pressure screw 6.2744.010

53 Capillary

PEEK capillary 6.1831.010 or PTFE-

microcapillary 6.1803.030

56 Filter unit PEEK 6.2821.120

50

53

56

50

53