7.1 Technical data

761 SD Compact IC / Instructions for Use 8.

761.1043

107

7 Appendix

7.1 Technical

data

Unless otherwise specified, the published data com-

prises typical values for the 761 SD Compact IC at an

ambient temperature of 25 °C.

7.1.1 Conductivity

measurement

Measurement range 1

0

1000

µ

S/cm (resolution: 0.56 nS/cm)

Measurement range 2

0

250

µ

S/cm (resolution: 0.14 nS/cm)

Measurement range 3

0

50

µ

S/cm (resolution: 0.028 nS/cm)

Maximum error

±

1 % of full scale value and

±

1 % of measurement value (k = 16.7/cm)

Linearity

Deviations <

±

0.5 % of full scale value

Noise

Measurement

range

1

Measurement

range

2

Measurement

range

3

typ. 10 nS/cm

typ. 2.5 nS/cm

typ. 0.5 nS/cm

Drift (electronic)

typ. < 10 ppm/h of full scale value

Temperature dependence

typ. < 40 ppm/°C of full scale value

Reserve range

> 33 % (k = 16.7/cm)

Sampling rate

10 measurements/s (fixed)

7.1.2 Conductivity

detector

Construction

Thermostatted conductivity detector with

2 ring-shaped steel electrodes

Measurement principle

Alternating current measurement with 1 kHz fre-

quency and ca. 1.7 V amplitude (peak to peak).

Cell volume

1.5

µ

L

Cell constant

approx. 17 /cm (the exact value is printed on the

detector)

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to adjustable oper-

ating temperature

Operating temperature

Adjustable in steps of 5°C from 25

45°C

Advertising