1 installing the mpak cabinet, 2 drain tube for inner compartment, 3 drain tube for mpak cabinet – Metrohm 761 SD Compact IC User Manual

Page 24: Installing the mpak cabinet, Drain tube for inner compartment, Drain tube for mpak cabinet

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

16

The inlet and outlet capillaries (

39

and

41

) to and from the detector

block must be fitted in the eluent path as described in Sections 2.6.5

and 2.6.6.

2.5

Installation of the MPak cabinet and connec-
tion of the drain tubes

2.5.1

Installing the MPak cabinet

Install the MPak cabinet as shown on the drawing accompanying the

package. Position it next to the 761 SD Compact IC and suspend the

eluent and regenerant MPaks from the crossbars. Since eluent and re-

generant are supplied from the rear to the 761 SD Compact IC, the out-

lets of the MPaks should point to the rear.

2.5.2

Drain tube for inner compartment

The 761 SD Compact IC features a connector (5) for escaped fluids on

the front panel. A drain tube can be fitted to the connector. Proceed as

follows:

1 Connect drain tube

Plug the silicone tube 6.1816.020 onto the nipple.

2 Route the drain tube into an outlet

Route the other end of the drain tube into an outlet and se-

cure it at this point.

2.5.3

Drain tube for MPak cabinet

The basin of the MPak cabinet 22 features a connection for fluids es-

caping from the MPaks to which a drain tube can be fitted. Proceed as

follows:

1 Connect drain tube

Plug the silicone tubing 6.1816.020 onto the nipple.

2 Route the drain tube into an outlet

Route the other end of the drain tube into an outlet and se-

cure it there.

Advertising